Technical article for PRP users

How to Set PRP Centrifugation Correctly: RCF, RPM, Rotor and Time

In professional PRP preparation, tubes, anticoagulant or separation gel, centrifuge, rotor, runtime and handling all work together. Misunderstandings most often occur with RPM, RCF and time.

This guide explains the difference between RPM and RCF, shows conversion using rotor radius and provides technical orientation for reproducible PRP and PRF preparation.

PRP centrifugationRCF RPM PRPSet PRP centrifugeCalculate G-force

RCF instead of RPM only

RCF describes the actual G-force acting on the sample and is more comparable across devices than RPM alone.

Consider rotor radius

The same RPM can generate different RCF depending on the rotor. The radius must therefore be considered.

Work system-specifically

Tube, rotor, RCF, time, brake setting and IFU must fit together.

RPM vs. RCF: why speed alone is not enough

RPM only describes how fast the centrifuge rotor rotates. For the sample, the actual relative centrifugal force is decisive: RCF or G-force.

Two centrifuges can run at the same RPM and still generate different forces because rotor radius differs. An instruction such as “3,000 RPM for 10 minutes” is therefore only transferable to a limited extent without rotor radius.

  • RPM
  • rotor radius
  • RCF / G-force
  • rotor type
  • runtime
  • tube and protocol

The RCF formula: converting RPM into actual G-force

Relative centrifugal force can be calculated with the following formula:

RCF = 1,118 × 10⁻⁵ × r × RPM²

RCF = relative centrifugal force in × g, r = rotor radius in centimetres, RPM = revolutions per minute. If a protocol specifies 1200 × g and the rotor radius is about 8.6 cm, the speed is approximately 3,500 RPM.

Practical note: calculate RCF/RPM directly with the tool

To avoid manual conversion, you will find a separate RCF/RPM conversion tool directly below this article. There you can enter the rotor radius of your centrifuge and calculate the suitable RPM for a desired RCF or G-force.

The tool is intended as mathematical orientation and does not replace the instructions for use of the medical device, the centrifuge manual or the professional assessment of the user.

Rotor type: fixed-angle and swing-out rotors

In addition to RPM and RCF, rotor type influences layer formation and handling after centrifugation. Fixed-angle and swing-out rotors produce different sedimentation geometries.

Neither rotor type is automatically better. The decisive point is whether rotor, tube, RCF, time and manufacturer protocol fit together.

  • fixed-angle rotor: compact and common in benchtop centrifuges
  • swing-out rotor: horizontal orientation during the run
  • adapters, buckets and tubes must match the device

Runtime, brake and soft stop: avoid unnecessary layer disturbance

Runtime should never be considered in isolation. A protocol with lower RCF and longer runtime can produce a different result from one with higher RCF and shorter runtime.

After centrifugation, blood components are separated into layers. If device and protocol allow it, gentler braking may help keep the layer image calm.

  • choose brake setting according to device manual
  • remove tubes calmly and vertically
  • avoid strong shaking
  • do not unnecessarily tilt or mix separation-gel systems

Single-spin or double-spin: decide system-specifically

With single-spin, preparation takes place in one centrifugation run. This is often used in defined PRP systems, especially with tubes containing anticoagulant and separation gel.

With double-spin, the sample is processed in two centrifugation steps. Whether this is useful depends on system, desired PRP type and manufacturer or protocol specification.

Example Vi PRP-PRO: think in RCF, not only RPM

Vi PRP-PRO is often used in practice with a defined single-spin protocol. The decisive value is not RPM alone, but the target RCF.

If the product protocol specifies 1200 × g for 7 minutes, the suitable RPM must be calculated using the rotor radius of the centrifuge.

Orientation values for PRP and PRF: use only system-specifically

The following information is not a general treatment recommendation. It only shows that PRP and PRF protocols can vary significantly depending on the system.

Protocol typeTypical viewImportant note
PRP with separation gel, e.g. Vi PRP-PROdefined single-spin according to product protocolRCF, time and tube must fit the manufacturer protocol
PRP without separation geloften system-dependent, sometimes single- or double-spinno blanket transfer of external RPM values
PRF / PRF-related protocolsoften without anticoagulant and without separation geltime-critical because clotting starts immediately
Double-spin PRPtwo centrifugation steps possibleuse only if protocol and product provide for it

Especially with PRF, tubes without anticoagulant must be processed immediately after blood collection. Time, temperature, tube material and centrifuge are closely linked to the protocol.

Temperature and device location: practical, but do not improvise

Centrifuges can warm up during longer use. Temperature conditions should not be improvised, but should follow the device manual, tube information and protocol.

Precooled adapters, opened lids or other measures should not be recommended generally. Safe and intended centrifuge operation remains decisive.

  • do not operate in direct sun or next to heat sources
  • ensure sufficient ventilation
  • observe manufacturer information on duty cycle
  • use adapters and buckets only as intended

Common mistakes in PRP centrifugation

Typical mistakes include copying RPM values without rotor radius, mixing PRP and PRF protocols or using unsuitable adapters and tubes.

Experience can be helpful, but it does not replace IFU, device manual and intended purpose of the product used.

  • copying RPM without rotor radius
  • mixing PRP and PRF protocols
  • not matching tube and centrifuge
  • replacing manufacturer information with experience

RCF / RPM

RCFRadiusRPM
1200 × g8.0 cmapprox. 3,660 RPM
1200 × g8.6 cmapprox. 3,530 RPM
1200 × g9.0 cmapprox. 3,450 RPM
1200 × g10.0 cmapprox. 3,275 RPM

Check question

Check questionWhy important?
Is the target RCF known?RCF is more comparable across devices than RPM.
Is the rotor radius known?Without radius, RPM cannot be correctly converted into RCF.
Does the tube fit the rotor?Length, diameter and adapters must be compatible.
Are tubes correctly balanced?Important for device safety and smooth running.
Is the brake setting suitable?Excessive braking can influence the layer image.
Is the IFU followed?Manufacturer information is decisive.

Conclusion: always consider RCF, RPM and time as one system

Professional PRP centrifugation is not just pressing a button. It is a standardised process in which tube, centrifuge, rotor, RCF, RPM, runtime and handling must fit together.

The most important step is moving from RPM-only thinking to understanding actual RCF or G-force. Only then can protocols be technically transferred between different centrifuges in a traceable way.

FAQ

FAQ: PRP centrifugation, RCF and RPM

What is the difference between RPM and RCF?

RPM describes rotor speed in revolutions per minute. RCF describes the relative centrifugal force, the actual G-force acting on the sample.

Why is RPM not enough for PRP?

An RPM value is not clear without rotor radius. Two centrifuges can create different RCF values at the same RPM.

How do you calculate RCF from RPM?

The formula is: RCF = 1.118 × 10⁻⁵ × r × RPM², where r is the rotor radius in centimetres.

Which RCF is suitable for Vi PRP-PRO?

The product protocol is decisive. If 1200 × g for 7 minutes is specified, the suitable RPM must be calculated using rotor radius.

Is a fixed-angle or swing-out rotor better for PRP?

This cannot be answered generally. Both can be suitable if centrifuge, tube, adapter, RCF, time and protocol fit together.

Can PRP protocols be transferred from one centrifuge to another?

Only with conversion and checking. RCF, rotor radius, rotor type, tube, filling volume, time and manufacturer information are decisive.

Note for professional users: This article provides technical information about the physical principles of centrifugation. The instructions for use of the medical devices used, the centrifuge manual, the manufacturer’s intended purpose and professional assessment remain decisive.

Lab calculator

PRP/PRF centrifugation calculator

Select a protocol or enter your own values to calculate RCF and RPM for PRP and PRF centrifugation protocols.

Important: RPM values cannot simply be transferred from one centrifuge to another. Rotor radius, RCF/×g, time, tube type and the validated protocol are decisive.

Step 1 — Select protocol

Selected: PRP-PRO. Target RCF 1200 ×g and 7 minutes have been applied.

Step 2 — Rotor radius

This means the distance from the center of rotation to the bottom of the tube, not the diameter. The preset value of 100 mm is only a reference value and must be checked against your own centrifuge.

Why the radius matters

The same RPM setting can produce a different RCF depending on the centrifuge. Example: 3500 RPM at an 86 mm radius gives a different ×g value than 3500 RPM at a 100 mm radius.

Step 3 — Calculation

Required speed — RPM

Formula used

RCF = 1.118 × r(mm) × (RPM / 1000)²

Rearranged: RPM = √(RCF / (1.118 × r(mm))) × 1000. The rotor type should be documented, but it is not calculated with a general correction factor.

Reference table

Protocol Typical value Time RCF at 100 mm Tube / material Classification
PRP-PRO Target: 1200 ×g 7 min approx. 3277 RPM at 100 mm PRP tube with anticoagulant and separating gel For PRP preparation with gel separator tubes
L-PRF / Choukroun 2700 RPM 12 min ≈815 ×g Glass / PRF tube Classic PRF protocol, radius-dependent
A-PRF 1500 RPM 14 min ≈252 ×g Glass / PRF tube Low-speed-oriented PRF protocol
A-PRF+ 1300 RPM 8 min ≈189 ×g Glass / PRF tube Modified low-speed PRF protocol
i-PRF 700 RPM 3 min ≈55 ×g Plastic tube without anticoagulant Liquid PRF fraction, time-critical processing
T-PRF 2800 RPM 12 min ≈877 ×g Titanium tube Titanium tube-based PRF protocol

The table shows reference values at a rotor radius of 100 mm. The actual RCF depends on the radius of the centrifuge used. RPM values from literature or manufacturer protocols should therefore not be transferred to other centrifuges without checking. The validated protocol, the instructions for use of the tube or centrifuge system and the professional responsibility of the treating facility remain decisive.

Related products

Centrifuge Hettich EBA 200 Centrifuge Hettich EBA 200 2
immediately available
Centrifuge

Hettich EBA 200 Centrifuge

HETTICH
100347
€1,235.29
The Hettich EBA 200 MD is the medical device version of the compact EBA 200 centrifuge and is designed as a Class IIa medical device for professional use in medical facilities. It is ideal for small sample volumes in daily practice, laboratory and clinical routines. Product features: MD version: Class IIa medical device Hettich article number: 1810...
Product added to wishlist
Product added to compare.
group_work Cookie consent