Matching PRP tubes and centrifuges correctly
Compatibility, RCF, RPM, rotor radius and a traceable PRP protocol – explained using the Vi PRP-PRO and Hettich EBA 200 MD reference system.
The correct setting starts with the tube
In PRP preparation, the tube, rotor, adapter, centrifugal force and run time form one connected system. A rotational speed can therefore never be assessed independently of the rotor radius.
The primary target specified by the tube manufacturer is the relative centrifugal force, abbreviated RCF or × g. The required speed for the specific centrifuge is calculated from this value using the rotor radius.
The key question is therefore not: “Which speed is best for PRP?” Instead: “What force and run time does the tube system require, and what speed produces that force in my centrifuge?”
A familiar RPM value is not a transferable PRP protocol. Without the rotor radius, it remains technically incomplete.
Same speed, different force
The same RPM value can produce a different relative centrifugal force with different rotors. The effective radius extends from the axis of rotation to the bottom of the tube in its centrifugation position.
With a smaller radius, the resulting RCF is lower. A familiar speed must not be adopted without checking.
As the distance from the axis of rotation increases, centrifugal force rises – even though the selected RPM remains unchanged.
A larger rotor produces more × g at the same speed. That is why the rotor radius is essential for every correct conversion.
RCF = 1.118 × 10⁻⁵ × r × RPM²
The radius r is entered in centimetres. For 1,200 × g at a radius of 8.6 cm, the calculated speed is approximately 3,530 rpm.
The reference system in detail
Vi PRP-PRO and Hettich EBA 200 MD are used here as a clearly defined combination. The values must not be transferred to other tubes or centrifuges without verification.
Vi PRP-PRO
Sterile borosilicate-glass vacuum tube with sodium citrate and a thixotropic separation gel.
Hettich EBA 200 MD
Compact centrifuge with a fixed 8-place angle rotor for medical facilities.
When are a tube and centrifuge compatible?
Dimensions and shape
The tube must sit fully, upright and without mechanical stress in the intended holder. Diameter, length and bottom shape are more important than nominal volume alone.
Permissible centrifugal force
The protocol’s target RCF must be achievable without exceeding the tube’s permissible load.
Effective rotor radius
The conversion requires the radius of the specific tube position – not the rotor diameter and not a value taken from another device.
Rotor and adapter
A different adapter or tube length changes the position of the tube bottom. This can also alter the calculated speed.
Symmetrical loading
Opposing positions are loaded with tubes of the same type and comparable mass. Imbalance monitoring does not replace correct balancing.
Reference protocol for Vi PRP-PRO and EBA 200 MD
This workflow describes technical preparation at system level. It is not a treatment instruction and does not replace the instructions for use or the clinical decision of the responsible professional.
| Parameter | Reference value | Why it is documented |
|---|---|---|
| Target RCF | 1,200 × g | Describes the relative centrifugal force actually applied. |
| Run time | 7 minutes | Forms part of the product-specific preparation workflow. |
| Rotor radius | 86 mm | Required for conversion between RCF and RPM. |
| Speed | approx. 3,500 rpm | Applies to the described reference configuration, not universally. |
What a PRP protocol should document
Traceable documentation makes troubleshooting easier and prevents an isolated speed value from being recorded without its system context.
What goes wrong when transferring settings
Using the RPM from another device
A different rotor radius can produce a substantially different RCF at the same speed.
Treating RCF and RPM as equivalent
1,200 × g is not 1,200 rpm. They are different physical quantities.
Using overly broad values
Broad ranges such as 6–10 minutes do not replace a product-specific protocol.
Underfilling the tube
In a vacuum tube, insufficient blood collection changes the ratio to the anticoagulant already present.
Balancing dissimilar masses
Opposing tubes should be comparable in design and filling level.
Mixing up PRP and PRF
Tubes with sodium citrate and separation gel follow a different system from conventional PRF tubes without anticoagulant.