The centrifuge is a central part of PRP and PRF preparation. Speed and price matter, but intended use, tube compatibility, RCF range, rotor, usability and documentation are just as important.
Check MDR or IVDR version, intended use and instructions for use.
RCF range, rotor radius, rotor angle, runtime and braking behaviour affect preparation.
Do tubes fit? Are settings easy? Is service available? Is cleaning practical?
PRP and PRF preparation is not just routine sample handling. The processed blood components become part of a medical or aesthetic-medical workflow. A device should therefore not be selected only by maximum speed or low purchase price.
A suitable centrifuge should enable reproducible preparation conditions: an appropriate rotor, matching adapters, stable RPM or RCF settings, defined runtimes and handling that can be documented in daily practice.
The key point is the manufacturer’s intended use, as described in labelling, instructions for use and technical documentation. A centrifuge may be intended as a laboratory or IVD device, or as a medical device for defined medical applications.
For PRP or PRF workflows in which processed blood components are applied again, the complete process chain should be checked: centrifuge, tubes, collection system, sterile handling, documentation and medical indication. MDR marking alone is not a legal guarantee, but it is an important element when the device is used according to its intended purpose.
In practice, the first question should not be only “is it fast enough?”, but “is this device intended for my workflow and does it fit the tubes I use?”
Many protocols state either RPM or RCF. RCF is usually more useful for comparison because it includes the rotor radius. Two centrifuges can generate different g-forces at the same RPM.
The basic formula is: RCF = 1.118 × 10-5 × radius in cm × RPM². A practical overview is available in the internal article PRP centrifugation: RCF, RPM, rotor and time.
The setting should always match the tube, system and protocol. For PRF, braking behaviour should also be considered.
The rotor affects how blood components separate during centrifugation. Fixed-angle rotors are compact, quick to use and common in small practice centrifuges.
Swing-out rotors can support horizontal layering but usually require more space and are often used in larger devices. The relevant question is not generally better or worse, but whether rotor angle, tube length, adapter and sediment position fit the protocol.
| Rotor type | Typical strength | What to check |
|---|---|---|
| Fixed-angle rotor | Compact, robust, integrated in many practice centrifuges | Check rotor angle, tube length, adapters and sediment position |
| Swing-out rotor | Horizontal layering during centrifugation | Consider size, cost, runtime, compatibility and maintenance |
PRP and PRF are often mentioned together but are prepared differently. PRP is usually collected with anticoagulant so the plasma remains liquid before application. PRF is typically prepared without anticoagulant, making the time between blood draw and centrifugation especially relevant.
PRF protocols may need lower g-forces, defined runtimes and adapted braking behaviour. PRP often focuses on clear separation, reproducible plasma yield and clean fractioning.
A protocol from another centrifuge should not be copied blindly. RPM values must be converted to RCF using rotor radius. Rotor angle, tube format and braking behaviour should also be considered.
The Hettich EBA 200 MD is a compact centrifuge for small sample volumes and is offered as an MDR version for medical applications. According to the datasheet, it is a medical device under Regulation (EU) 2017/745 for separating whole blood or blood components of human origin.
Technically, the EBA 200 MD has an integrated 8-place fixed-angle rotor, a speed range up to 6,000 min-1, a maximum RCF of 3,461 x g and a maximum capacity of 8 x 10 ml. Before purchase, tube format, protocol and intended application should still be checked.
| Criterion | Practice assessment |
|---|---|
| Regulatory status | MDR version available; intended use and IFU must fit the application |
| Rotor | integrated 8-place fixed-angle rotor |
| RCF/RPM | up to 3,461 x g or 6,000 min-1 |
| Tubes | tube format, diameter, length and adapters must be checked in advance |
| Practice use | compact, straightforward operation, suitable for small sample volumes |
Many wrong purchases happen not because the device is bad, but because it does not fit the planned process.
For PRP and PRF, the centrifuge is central, but it never works in isolation. The relevant system includes centrifuge, rotor, tubes, protocol, sterile workflow and documentation.
A compact MDR centrifuge such as the Hettich EBA 200 MD can be a suitable option for many practices if the chosen tubes and protocols are compatible. The decision should not be based on price or one technical value alone.
Professionals can find suitable devices in the centrifuge category. The product page for the Hettich EBA 200 MD provides further product-related information. Suitable PRP tubes are also relevant for preparation. This reference concerns material, device selection and workflow organisation, not clinical efficacy.
This article is intended for healthcare professionals and provides general information on device selection. It does not replace checking intended use, instructions for use, local regulatory requirements, hygiene rules or the professional responsibility of the practitioner.